| |
|
|
Glossary of
Electrosurgery
|
|
|
Active (electrode)
|
Electrode
that can be connected to the electrodes-holder handle thanks to that
it is possible to transfer the HF output power to the patient,
therefore, to obtain the selected performance.
This electrode is used for monopolar application of electrosurgical
units together with neutral electrode.
|
|
Argon
|
Inert
gas used by special units (called “Argon coagulators”) for obtaining
a coagulating effect on the patient tissue. The gas inside the tank is delivered at a low pressure towards
the tissue while activating a function with high voltage, which brings
the Argon’s spark, then starting the coagulating process.
It's a further variant from the coagulating techniques.
|
|
Automatic Bipolar Coagulation
|
It
is not necessary to activate bipolar coagulation by the operator;
the contact between the forceps and the patient tissue activates the
unit and, thus, enables the coagulation.
|
|
Bi-coagulation
|
See
“Bipolar coagulation”.
|
|
Bipolar Coagulation
|
Type
of coagulation which uses the bipolar forceps that permits to deliver
the HF output power on an impedance having value of approximately
100 ohm. Such value is equal
to the one of the tissue section which is normally between the tips
of the forceps.
|
|
Bipolar (electrode)
|
Forcep electrode
thanks to that it is possible to utilize the bipolar performances of electrosurgical
units, i.e. those in which every electrode tip represents one of two poles
of the circuit where the patient in included.
|
|
Bipolar (operation)
|
The
way to use electrosurgical units being possible thanks to forcep electrodes,
where the HF output power is delivered between the 2 forcep tips,
thanks to that the surgeon
is able to carry out small coagulations.
|
|
Blend Cut
|
Waveform with a
percentage of modulation less than the pure coagulation. The modulation
is a gap thanks to that it is possible to deliver suitable “train impulse”
of HF energy. The width of the gap compared to the HF power delivery span
lets the water contained inside the cells to evaporate more or less quickly.
This comes to a coagulating effect more or less significant (haemostatic
effect).
|
|
Coagulation
|
Output waveform
that makes the water inside the cell to evaporate gradually, without making
the cell tissue bursted (as it happens when cutting).
|
|
Cut
|
Particular waveform that causes a
very quick evaporation of the water inside the cells of the patient
tissues; at macroscopic level it is a dissection of the tissue itself.
|
|
Defibrillator
|
Device,
compulsory by law in the operating theatre, able to supply a strong
electric discharge to patients in ventricular fibrillation condition.
It is used in order to recover the heart function that has suddenly
stopped beating. On the electrosurgical units it must be clearly stated
the protection from defibrillator discharge since, in case of emergency,
it could not be able to take off the neutral electrode from the patient.
In such cases the discharge does not have to create risks either for
the patient or the instrumentation connected to him.
|
|
Earth-referred (unit)
|
Is
defined as “earth-referred” an electrosurgical unit whose patient
circuit is electrically separated from the protective earth
or earthing equalizationsystem
only as regarding the low frequency currents. It means that,
when the device delivers the HF output power, the neutral electrode
(and thus the patient) is connected to earth. The electrical separation
between the patient circuit and the
earth is normally done by the use of a capacitor that blocks the low
frequency components (for which it acts as an high-valued impedance),
while it becomes almost “transparent” regarding the radiofrequency
output current of the device itself.
|
|
Electrode-holder pencil
|
Plastic
accessory, with or without buttons on its body depending on the type
of activation, that permits using the monopolar electrodes.
|
|
Electromagnetic compatibility
|
Characteristic that permits
the electric units to interact with others without undergoing or causing
such interferences that can compromisetheir correct operation. Some tests able to verify the electromagnetic emission level
and their immunity have been carried out to verify such compatibility.
|
|
Floating (unit)
|
“Floating”
is used for a surgical unit whose patient circuit is always found
separated from the protective earth.
These are units that can guarantee the highest level of safety
from the dangers relevant to the dispersion currents.
|
|
Fulguration
|
Particular
type of coagulation that
permits to cauterize small blood vessels. A difference from the normal coagulations by contact,
the Fulguration
allows the coagulation by starting a spark that produces the desired
effect with particular precision and speed, thanks to using a high
voltage of output waveform.
|
|
HF Leakage current
|
High
Frequency current whose effect can cause a burn in the patient.
It's one of the main problems using an eletrosurgical unit
It is normally caused by a coupling between the patient and the earth,
due to an accidental contact of the patient himself with a conductive
part connected to earth, or by using of other devices for monitoring
during the operation.
|
|
Laparoscopy
|
Surgical technique that
allows to operate the patient inside him abdomen.
The operating field remains closed and it is possible to enter
in the zone subjected for the operation thanks to three small probes
(one of them is used for introducing the camera, the others for the
accessories connected to the electrosurgical units) inserted in the
abdomen of the patient. Where
is possible to apply this technique, the post-operation recover time
reduces remarkably, since the trauma due to the wound is
minimized.
|
|
LF Leakage current
|
Electric
current, normally at mains frequency , that must be absolutely kept
within the limit prescribed by the general standard for the safety
of the medical units CEI EN 60601-1. The observance of the imposed values avoids the risk of fulguration
for the patient or the operator.
|
|
Monopolar (operation)
|
Way
of using the electrosurgical unit by the active electrode and neutral
electrode; with having
these 2 electrodes, the active one is used for cutting and coagulating
the patient tissue because of its very reduced surface, while the neutral one acts exclusively
as return electrode.
|
|
Neuromuscular Stimolation
|
Effect
caused by the low frequency alternating current. The central nervous system works thanks to the delivering low
frequency electric pulses. A
low frequency current stimulates the muscles and causes their contraction.
This happens also for the cardiac muscle, that can go in “fibrillation”.
To avoid such dangerous effect the frequency of the current
delivered by an electrosurgical unit must be high enough in order
not to allow the nerves of being able to “react”.
|
|
Neutral (electrode)
|
Flat
electrode of surface approximately 150-200 cm2 essential in monopolar working of the electrosurgical unit,
that allows the “return” of the current (which has produced the cutting or the coagulation)
towards the device.
|
|
Neutral plate
|
See “Neutral
(electrode)”.
|
|
Radiofrequency
|
Range
of requency (normally from 300 kHz up to 3 MHz) within that it is
selected the operative frequency of an electrosurgical unit. Values
under 300 kHz could increase the risk of neuromuscolar stimulation
for the patient; values above 3 MHz
could create difficulties in control the high frequency dispersion
currents (and therefore could increase the burning risk on the patient).
|
|
Rated load
|
The value of non-reactive
load resistance that results in the maximum HF output power from each
operating mode of the HF surgical equipment.
|
|
Spray Coagulation
|
Particular
kind of coagulation that takes place without contact between the active
electrode and the patient tissue. The output voltage is even higher than the Fulguration and
produces an effect similar to a spray of a canister (the name of “spray”
comes from that).
|
|
T.U.R.
|
Is the acronym
for “Trans-Urethral Resection”.
|
|
|
|