SMOKE EVACUATORS
CS
SERIES
|

|
The smoke evacuators CSU are innovative suction units, with new
characteristics, which have been designed to evacuate and filter the
smoke produced during all the electro-surgical procedures, by eliminating the bad smells or
the bacteriological risks,
and by solving the very common problem
of the visual field during the laparoscopic procedures.
The CSU200 is equipped with an electric suction
group at high flow and medium noise, moreover it has got the
function TURBO - high suction. It represents the basic model, which
is ideal for the out-patient treatment, thanks to the connection
with the speculum in gynaecology, the suction systems on the
electrodes-handle, and the fixed tubing on the operation
point.
The CSU900LC - the most complete model - is
equipped with an electric suction group at high flow and low noise,
a special circuit for the connection to a centralized vacuum system,
and another circuit for the controlled
output of the CO2 which is used in the
laparoscopic procedures with the pneumoperitoneum. It is the ideal
model for the usage in the operating theatre, thanks to the low
level of noise produced, and the big range of the available
procedures. It allows what follows:
- the evacuation of the smoke during the surgical open
sky procedures, thanks to the aspiration system on the
electrodes-handle or by an
independent cannulae, trough the suction group or the external vacuum
system;
- the adjustable evacuation of the smoke during the
laparoscopic procedures with the CO2 pneumoperitoneum,
thanks to the connection with the trocar, through the controlled output of the gas itself;
- the adjustable evacuation of the smoke during the
gas-less laparoscopic procedures, thanks to the connection with the
trocar, through the suction group or the external vacuum
system.
The CSU200LC is the same as the previous one, but
it is equipped with a less powerful electric suction group. It is
ideal for the suction during the laparoscopic procedures. It allows
what follows:
- the evacuation of the smoke during the laparoscopic
procedures with the CO2
pneumoperitoneum, through the adjustable leakage of the
gas itself;
- the evacuation of the smoke, with adjustable flow,
during the gas-less laparoscopic procedures, through the suction
group or the external vacuum
system.
All the models can be used as a “shelf” for the electro
surgical units; there are also specific trolleys, both for the
out-patient treatment and the operating
theatre.
|
|
TECHNICAL
FEATURES |
|
GENERAL
TECHNICAL FEATURES |
|
Microprocessor
Control System - Functioning control through micro-processor,
with auto-check at the switching-on
Three
Stages Filter -
Three stages filter: two antibacterial Hepa type, with a total
efficiency 99,99999% for
Ć
0. 3 micron particles, and one anti-smell with active carbons,
controlled by a micro-processor with double check (filter efficacy,
limit time of usage of 2 hours), luminous alarm signal and
functioning block after the first switching-on before the intervention of the control
circuit
Automatic
Start/Stop Sensing System - Automatic
switching on / switching off of the suction system, every time the
electro surgical unit is activated, through a non-contact sensing system which
eliminates the need of both
specific connections to the electro surgical unit and a pedal
switch (available just in case the operator prefers this
option)
Programmable
Stop Delay - Adjusting of the delay in the
switching off of the suction system from 0 to 30sec, in order to
completely eliminate the smoke
Switching
on Pedal -
Pneumatic, waterproof, anti-explosion, for operating theatre
IPX7/AP/APG (non standard, but available on
request)
Enclosure
- Metallic
and painted, it can be used as a “shelf” for the electro surgical
unit, or it can be put in specific
trolleys
Cooling
–
By convection, no fan
Classification
IEC 60601-1 -
Class I, type BF (leakage to the enclosure: less than 25microA;
leakage to the patient: absent)
Safety
Rules
- The units meet IEC 601-1 Safety Rules
Classification
and Approval EC 93/42 -
Class IIA -
CE0051
Dimensions
and Weight
-
40x40x21cm,
17kg
|
|
|
|
TECHNICAL
FEATURES CS200 |
|
Suction
Group -
Rotative, non lubricated - Flow: 500lt/min; Max. vacuum:
110mmHg
Suction
Regulation -Electronic,
with function Turbo - high suction
Rotative
control with visual bar-led
Mains
/ Absorption / Fuses / Switch
- 230V-50/60Hz
/ 650VA / 2x5A / Bipolar,
green. |
|
|
TECHNICAL
FEATURES CS900LC |
|
Suction
Group - Rotative,
lubricated - Flow:50lt/min; Max. vacuum:
720/730mmHg
Suction
Regulation - Rotative control, with
graduated scale
Mains
/ Absorption / Fuses / Switch
-
230V-50/60Hz / 250VA / 2x1,6A / Bipolar,
green
Connection
to the external vacuum system - Connector with rapid
device
Control
of the CO2 output
-
Circuit with pressure tight
electro-valve. |
|
|
TECHNICAL
FEATURES CS200LC |
|
Suction
Group -
Rotative, with membrane / piston - Flow: 23lt/min; Max. vacuum:
650/700mmHg
Suction
Regulation - Rotative
control, with graduated
scale
Mains
/ Absorption / Fuses / Switch - 230V-50/60Hz
/ 138VA / 2x1A / Bipolar, green
Connection
to the external vacuum system - Connector
with rapid device
Control
of the CO2
output
-
Circuit with pressure tight
electro-valve. |
|